Package gda.px
Class MxProperties
java.lang.Object
gda.px.MxProperties
Constants for MX property names.
-
Field Summary
FieldsModifier and TypeFieldDescriptionstatic final booleanFlag to autoprocess incrementally for every run of a data collection group.static final Stringstatic final StringFormat string for the number component of a folder name.static final intstatic final StringThe Geobrick anneal out has no state because it has no limit switchstatic final StringBoolean of whether to show the panel created by SharedGuis.createAnnealPanel.static final Stringstatic final StringName of Scannable with position of value "active" when the further actions that effect the hutch are to be disabled - hutch utils script is running or sampleChangerState is Active or a data collection is being performed as indicated by GDA_DATACOLLECTION_SCRIPTstatic final StringName of file into which chooch results are to be writtenstatic final Stringused for clientSampleMovementService, gonioOnLeftOfImage fieldstatic final StringName of Scannable with position equals the name of the active data collection script - or else ""static final StringBoolean of whether the message "NOTE: backstop must be in beam before centring" will be displayed.static final StringMax image time in table.static final Stringstatic final StringMax omega delta.static final StringMaximum omega angle (degrees).static final StringMax sample step in table.static final StringMinimum image time in table.static final Stringstatic final StringMinimum omega delta.static final StringMinimum omega angle (degrees).static final StringMinimum sample step in table.static final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final StringTell the user that the beamline does not use attenuation optimization for fluorescence scan, that a spectrum/MCA scan will have to be donestatic final Stringstatic final StringName of file into which fluorescence spectrum results are to be writtenstatic final StringName of panel to which fluorescence spectrum results are to be plottedstatic final Stringstatic final Stringstatic final StringName of Scannable.static final StringWhen set totrue, prevents the beam centre from being set.static final StringWhen set totrue, prevents the beam size from being set.static final StringAddress of the sample display RTP stream.static final StringPort number of the sample display RTP stream.static final StringWhether the beam size boxes shown in the "Sample Alignment and Centring" tab are editable.static final StringIf set totrue, the beam size stored in the file is used instead of the real beam size.static final StringWhether the slit gap boxes shown in the "Sample Alignment and Centring" tab are editable.static final Stringstatic final StringDuring anneal, change the number of times to do the timeout check when the cryojet fine is moved outstatic final Stringstatic final Stringstatic final StringThe field name of the change filters apply PV.static final StringBoolean of whether Phase I-type of array of filters are used for attenuation.static final StringBoolean of whether attenuation filters need to be applied for the changes to be propagated to the hardware.static final StringWhether to show the beam size panel in the image control panel.static final Stringstatic final Stringstatic final StringName used in call to LocalParameters.getNewXMLConfiguration to open a XML Configuration file for beamline parameters Default value is mx_parametersstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final StringWhether to show the 'show beam scale' checkbox in the image control panel.static final Stringstatic final Stringstatic final doublestatic final Stringstatic final Stringstatic final StringControls whether values shown in the dashboard are or editable or not.static final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final StringMethod for calculating target dose or exposure.static final Stringstatic final Stringstatic final StringTrue if data collection form to show option to set a target dosestatic final StringThe name of the file storing a list of edges acceptable for fluo scans.static final StringThe big gonio nudge, in degrees (e.g.static final StringThe little gonio nudge, in degrees (e.g.static final StringName of the scannable that represents the omega axis of the goniometer.static final StringSuffix of epics motor record for gonio rotation axis.static final StringDefault value for GDA_MX_GONIO_ROTATION_AXIS_RECORDstatic final StringIf the HC1 device is ready and available for use in experimentsstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final intstatic final StringIdentifies the maximum number of valid locations for sample holders (eg puck locations in dewar)static final StringWhen set totrue, middle-clicking on the sample image causes the coordinates to be displayed.static final Stringstatic final Stringstatic final StringIdentifies the minimum valid integer index for numbered sample holders (eg first puck location in dewar)static final Stringstatic final Stringstatic final StringUsed for Message of the Day (Messages > Beamline notes)static final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final StringWhether pin tip centring is enabledstatic final Stringstatic final Stringstatic final StringWhether the rotation axis alignment panel is enabled.static final StringSpecifies whether the sample changer supports position blocking.static final StringName of the function in the sample changer script for loading samples.static final StringName of the sample changer script controller.static final StringName of the function in the sample changer script for unloading samples.static final StringTrue if a sample changer is in usestatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final StringDefault screening values.static final Stringstatic final StringShow Barcode in MX Controlsstatic final Stringstatic final Stringstatic final StringTrue if data collection form toExperiment Type selectionstatic final Stringstatic final Stringstatic final StringWhether OAV camera zoom controls use +/- buttons ( otherwise Combo )static final Stringstatic final StringShow the slit gap and controls for showing slit gap in GUIstatic final Stringstatic final StringWhether the wavelength column should be shown in the data collection table.static final StringInclude sample centring X/Y/Z positions in data collection parametersstatic final StringInclude V2 sample centring X/Y/Z position tools in data collection GUI (I23-1191)static final StringIf the users are to be given a choice of which axis to oscillate: Phi or Omegastatic final Stringstatic final Stringstatic final StringIf set to false then the columns related to data folder will be hidden in the data collection tablestatic final StringRather than having both kappa and phi displayed by the GDA_MX_MINIKAPPA_AVAILABLE property, only show kappastatic final StringIf the detector is mounted on a two theta armstatic final Stringstatic final Stringstatic final StringName of the scannable that represents the horizontal beam size.static final StringName of the scannable that represents the vertical beam size.static final Stringstatic final StringFolder containing Extended Collect Request xml files.static final StringWhich type of strategies to usestatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final Stringstatic final StringIf the sample changer is available, add _MxSy_ to the prefixstatic final StringAssumed to be true, but if set to false then parts of the UI relating to sample ids etc.static final Stringstatic final Stringstatic final StringDefault value for exposure time in experiment table.static final Stringstatic final Stringstatic final StringDefault value for omega oscillation in experiment table.static final StringDefault value for sample-detector distance in the experiment table.static final Stringstatic final Stringstatic final Stringstatic final StringDefault value for wavelength in experiment table.static final StringCommand to run at the end of the entire collection.static final StringCommand to run at the start of the entire collection.static final Stringstatic final Stringstatic final StringWhether there is an alternative sample stage scannable.static final StringInitial grid scan transmission for MX, in percentstatic final StringAllow user to use columns while setting up a grid scanstatic final StringAllow user to use rows while setting up a grid scanstatic final StringShould the GUI allow grid scan boxes to be selected for data collections?static final Stringstatic final StringAllow fast grid scans using the fast grid scan PMAC programstatic final StringShould fluo data be taken during fast grid scans?static final Stringstatic final StringInitial line scan exposure time for MX, in secondsstatic final StringInitial number of images per wedge for line scansstatic final StringInitial line scan oscillation for MX, in degreesstatic final StringInitial line scan resolution for MX, in ångströmstatic final StringInitial line scan transmission for MX, in percentstatic final StringInitial number of wedges for line scansstatic final StringWhether auto loop centring using the OAV (and sometimes Keyence) camera is enabled.static final Stringstatic final Stringstatic final Stringtrue if the sample changer is present.static final StringInitial still exposure time for MX, in seconds.static final StringMaximum still exposure time for MX, in seconds.static final StringMinimum still exposure time for MX, in seconds.static final StringWhether x-ray centring has been implemented on this beamline.static final StringName of Scannable with position equal to status of the sample changerstatic final StringName of Scannable with position of value that indicates the state of the sample control systemstatic final StringName of Scannable whose position indicates the current action of the sample control scriptstatic final StringName of Scannable.static final Stringstatic final Stringstatic final StringFormat string for the image number component of a filename template.static final Stringstatic final intMaximum possible image numberstatic final intFormat string for the number component of a folder name.static final intNumber of digits to use for the image number in filenames.static final intNumber of digits to use for the run number in filenames.static final Stringstatic final intstatic final booleanFlag to force ascending run numbers in a collection series when using manually specified run numbers I23-866static final StringFormat string for the run number component of a filename template. -
Method Summary
Modifier and TypeMethodDescriptionstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic doublestatic intgetMaxPrefixLength(boolean usingEiger) static intstatic intstatic intstatic Stringstatic doublestatic intstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic booleanstatic boolean
-
Field Details
-
GDA_PX_SAMPLE_CHANGER_PRESENT
true if the sample changer is present. This causes: SampleChanger to be added as a tab to JCameramanDisplayPanel Columns for holder and position are added to ExperimentControlPanel- See Also:
-
GDA_MX_SAMPLE_CHANGER_BLOCKED_POSITIONS_ENABLED
Specifies whether the sample changer supports position blocking. Requires a client-side blockedSamplePositions bean.- See Also:
-
GDA_MX_MAX_SAMPLE_HOLDER
Identifies the maximum number of valid locations for sample holders (eg puck locations in dewar)- See Also:
-
GDA_MX_MIN_SAMPLE_HOLDER_INDEX
Identifies the minimum valid integer index for numbered sample holders (eg first puck location in dewar)- See Also:
-
GDA_IMAGES_PREVENT_SET_BEAM_CENTRE
When set totrue, prevents the beam centre from being set.- See Also:
-
GDA_IMAGES_PREVENT_SET_BEAM_SIZE
When set totrue, prevents the beam size from being set.- See Also:
-
GDA_MX_GONIO_OMEGA_SCANNABLE
Name of the scannable that represents the omega axis of the goniometer.- See Also:
-
GDA_MX_SLIT_SIZE_X_SCANNABLE
Name of the scannable that represents the horizontal beam size.- See Also:
-
GDA_MX_SLIT_SIZE_Y_SCANNABLE
Name of the scannable that represents the vertical beam size.- See Also:
-
GDA_MX_BEAM_SIZE_X_SCANNABLE
- See Also:
-
GDA_MX_BEAM_SIZE_Y_SCANNABLE
- See Also:
-
GDA_MX_MIDDLE_CLICK_SHOW_COORDINATES
When set totrue, middle-clicking on the sample image causes the coordinates to be displayed.- See Also:
-
GDA_MX_SAMPLE_CHANGER_SCRIPT_CONTROLLER
Name of the sample changer script controller.- See Also:
-
GDA_MX_SAMPLE_CHANGER_LOAD_SCRIPT
Name of the function in the sample changer script for loading samples.- See Also:
-
GDA_MX_SAMPLE_CHANGER_UNLOAD_SCRIPT
Name of the function in the sample changer script for unloading samples.- See Also:
-
GDA_JCAMERAMANDISPLAYPANEL_SHOW_BEAM_SIZE_FROM_FILE
If set totrue, the beam size stored in the file is used instead of the real beam size.- See Also:
-
GDA_IMAGES_RTP_ADDRESS
Address of the sample display RTP stream.- See Also:
-
GDA_IMAGES_RTP_PORT
Port number of the sample display RTP stream.- See Also:
-
GDA_MX_CAMERA_SHOW_SCALE
Whether to show the 'show beam scale' checkbox in the image control panel.- See Also:
-
GDA_MX_SHOW_BEAM_CENTRE_CHECKBOX
- See Also:
-
GDA_MX_SHOW_FOCAL_SPOT_AS_ELLIPSE_CHECKBOX
- See Also:
-
GDA_MX_SHOW_MOVE_ON_CLICK_CHECKBOX
- See Also:
-
GDA_MX_SHOW_BARCODE
Show Barcode in MX Controls- See Also:
-
GDA_MX_STRATEGIES
Which type of strategies to use- See Also:
-
GDA_MX_STRATEGIES_COMMON
- See Also:
-
GDA_MX_STRATEGIES_ASTRA
- See Also:
-
GDA_MX_STRATEGIES_FILE
- See Also:
-
GDA_MX_BEAM_SIZE_PANEL_ENABLED
Whether to show the beam size panel in the image control panel.- See Also:
-
GDA_JCAMERAMANDISPLAYPANEL_BEAM_SIZE_IS_EDITABLE
Whether the beam size boxes shown in the "Sample Alignment and Centring" tab are editable.- See Also:
-
GDA_JCAMERAMANDISPLAYPANEL_SLIT_GAP_IS_EDITABLE
Whether the slit gap boxes shown in the "Sample Alignment and Centring" tab are editable.- See Also:
-
GDA_PX_FRAME_SAMPLEXYZ_HAVE_CHOICE
Whether there is an alternative sample stage scannable.- See Also:
-
GDA_MX_ROTATION_AXIS_ALIGNMENT_PANEL_ENABLED
Whether the rotation axis alignment panel is enabled.- See Also:
-
GDA_MX_COORD_SYS_CENTRING_PAUSE_SECONDS
- See Also:
-
GDA_MX_COORD_SYS_CENTRING_PAUSE_DEFAULT
public static final double GDA_MX_COORD_SYS_CENTRING_PAUSE_DEFAULT- See Also:
-
GDA_MX_GONIO_LITTLE_NUDGE
The little gonio nudge, in degrees (e.g. 90).- See Also:
-
GDA_MX_GONIO_BIG_NUDGE
The big gonio nudge, in degrees (e.g. 180).- See Also:
-
GDA_PX_STILL_EXPOSURE_TIME_MIN
Minimum still exposure time for MX, in seconds.- See Also:
-
GDA_PX_STILL_EXPOSURE_TIME_MAX
Maximum still exposure time for MX, in seconds.- See Also:
-
GDA_PX_STILL_EXPOSURE_TIME_INITIAL
Initial still exposure time for MX, in seconds.- See Also:
-
GDA_PX_GRID_SCAN_TRANSMISSION_INITIAL
Initial grid scan transmission for MX, in percent- See Also:
-
GDA_PX_LINE_SCAN_OSCILLATION_INITIAL
Initial line scan oscillation for MX, in degrees- See Also:
-
GDA_PX_LINE_SCAN_EXPOSURE_TIME_INITIAL
Initial line scan exposure time for MX, in seconds- See Also:
-
GDA_PX_LINE_SCAN_TRANSMISSION_INITIAL
Initial line scan transmission for MX, in percent- See Also:
-
GDA_PX_LINE_SCAN_RESOLUTION_INITIAL
Initial line scan resolution for MX, in ångström- See Also:
-
GDA_PX_LINE_SCAN_WEDGES_INITIAL
Initial number of wedges for line scans- See Also:
-
GDA_PX_LINE_SCAN_IMAGES_PER_WEDGE_INITIAL
Initial number of images per wedge for line scans- See Also:
-
GDA_EXPT_TABLE_MODEL_MIN_IMAGE_TIME
Minimum image time in table.- See Also:
-
GDA_EXPT_TABLE_MODEL_MAX_IMAGE_TIME
Max image time in table.- See Also:
-
GDA_EXPT_TABLE_MODEL_MIN_SAMPLE_STEP
Minimum sample step in table.- See Also:
-
GDA_EXPT_TABLE_MODEL_MAX_SAMPLE_STEP
Max sample step in table.- See Also:
-
GDA_EXPT_TABLE_MODEL_MIN_SAMPLE_OMEGA_START
Minimum omega angle (degrees).- See Also:
-
GDA_EXPT_TABLE_MODEL_MAX_SAMPLE_OMEGA_START
Maximum omega angle (degrees).- See Also:
-
GDA_EXPT_TABLE_MODEL_MIN_OMEGA_DELTA
Minimum omega delta.- See Also:
-
GDA_EXPT_TABLE_MODEL_MAX_OMEGA_DELTA
Max omega delta.- See Also:
-
GDA_EXPT_TABLE_MODEL_MIN_NUM_IMAGES
- See Also:
-
GDA_EXPT_TABLE_MODEL_MAX_NUM_IMAGES
- See Also:
-
GDA_PX_EXTENDED_COLLECT_REQUEST_PROCESSOR_START_COLLECTION_COMMAND
Command to run at the start of the entire collection.- See Also:
-
GDA_PX_EXTENDED_COLLECT_REQUEST_PROCESSOR_END_COLLECTION_COMMAND
Command to run at the end of the entire collection.- See Also:
-
GDA_SAMPLECONTROLSCRIPT_STATE
Name of Scannable. When non-zero the GDA_SAMPLECONTROL_STATE is set to active.- See Also:
-
GDA_SAMPLECONTROLSCRIPT_ACTION
Name of Scannable whose position indicates the current action of the sample control script- See Also:
-
GDA_HUTCHUTILSSCRIPT_STATE
Name of Scannable. When non-zero the GDA_HUTCHUTILS_STATE is set to active.- See Also:
-
GDA_BEAMLINEBUSY_STATE
Name of Scannable with position of value "active" when the further actions that effect the hutch are to be disabled - hutch utils script is running or sampleChangerState is Active or a data collection is being performed as indicated by GDA_DATACOLLECTION_SCRIPT- See Also:
-
GDA_SAMPLECHANGER_STATE
Name of Scannable with position equal to status of the sample changer- See Also:
-
GDA_SAMPLECONTROL_STATE
Name of Scannable with position of value that indicates the state of the sample control system- See Also:
-
GDA_DATACOLLECTION_SCRIPT
Name of Scannable with position equals the name of the active data collection script - or else ""- See Also:
-
GDA_MX_DASHBOARD_EDITABLE
Controls whether values shown in the dashboard are or editable or not. Should betrueorfalse.- See Also:
-
GDA_CHOOCH_RESULTS_FILE
Name of file into which chooch results are to be written- See Also:
-
GDA_FLUORESCENCE_SPECTRUM_RESULTS_FILE
Name of file into which fluorescence spectrum results are to be written- See Also:
-
GDA_FLUORESCENCE_SPECTRUM_RESULTS_PLOTNAME
Name of panel to which fluorescence spectrum results are to be plotted- See Also:
-
GDA_FLUORESCENCE_SPECTRUM_NO_ATTENUATION_OPTIMIZATION
- See Also:
-
GDA_FLUORESCENCE_SCAN_NO_ATTENUATION_OPTIMIZATION
Tell the user that the beamline does not use attenuation optimization for fluorescence scan, that a spectrum/MCA scan will have to be done- See Also:
-
GDA_FLUORESCENCE_SCAN_FORM_SHOW_TRANSMISSION_CONTROL
- See Also:
-
GDA_FLUORESCENCE_DEFAULT_DETECTOR
- See Also:
-
GDA_FLUORESCENCE_MCA_AVAILABLE
- See Also:
-
GDA_FLUORESCENCE_TCA_AVAILABLE
- See Also:
-
GDA_FLUORESCENCE_MIN_ACQUISITION_TIME
- See Also:
-
GDA_FLUORESCENCE_MAX_ACQUISITION_TIME
- See Also:
-
GDA_MX_DOSE_CALCULATION_METHOD
Method for calculating target dose or exposure.- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_WAVELENGTH
Default value for wavelength in experiment table. In not defined use 0.9795- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_SDDIST
Default value for sample-detector distance in the experiment table. In not defined use 289.3.- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_SAMPLE_STEP
Default value for omega oscillation in experiment table. In not defined use 1.0.- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_IMAGE_TIME
Default value for exposure time in experiment table. In not defined use 1.0.- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_OMEGA_START
- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_NUMBER_IMAGES
- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_TRANSMISSION_PC
- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_USE_TARGET_DOSE
- See Also:
-
GDA_PX_EXPTTABLE_DEFAULT_USE_TARGET_DOSE_BY_RESOLUTION
- See Also:
-
GDA_MX_BEAMLINEPARAMETERS_XML
Name used in call to LocalParameters.getNewXMLConfiguration to open a XML Configuration file for beamline parameters Default value is mx_parameters- See Also:
-
GDA_MX_BEAMLINEPARAMETERS_XML_DEFAULT
- See Also:
-
GDA_MX_THREE_CLICK_CENTRING_VISIBLE
- See Also:
-
GDA_MX_THREE_CLICK_CENTRING_WAIT_FOR_MOVES_TO_FINISH
- See Also:
-
GDA_MX_RADDOSE_AVAILABLE
- See Also:
-
GDA_PX_LOOP_CENTRING_ENABLED
Whether auto loop centring using the OAV (and sometimes Keyence) camera is enabled.- See Also:
-
GDA_MX_LOOP_CENTRING_OMEGA_STEPS
public static final int GDA_MX_LOOP_CENTRING_OMEGA_STEPS -
GDA_MX_SHOW_OAV_ZOOM_BUTTONS
Whether OAV camera zoom controls use +/- buttons ( otherwise Combo )- See Also:
-
GDA_MX_PIN_TIP_CENTRING_ENABLED
Whether pin tip centring is enabled- See Also:
-
GDA_PX_XRAY_CENTRING_ENABLED
Whether x-ray centring has been implemented on this beamline.- See Also:
-
GDA_PX_MANUAL_XRAY_CENTRING_ENABLED
- See Also:
-
GDA_PX_DEFAULT_CENTRING_OPTION
- See Also:
-
GDA_MX_ATTENUATION_FILTER_ARRAY
Boolean of whether Phase I-type of array of filters are used for attenuation. I24 does not- See Also:
-
GDA_MX_ATTENUATION_NEED_APPLY
Boolean of whether attenuation filters need to be applied for the changes to be propagated to the hardware. Phase I do, I04-1, I24 do not- See Also:
-
GDA_MX_ATTENUATION_APPLY_FIELD_NAME
The field name of the change filters apply PV. Phase I has ".PROC", I04 has ""- See Also:
-
GDA_DISPLAY_3CLICK_BEAMSTOP_WARNING
Boolean of whether the message "NOTE: backstop must be in beam before centring" will be displayed. This message was for Phase I where the beamstop used to affect the visual system- See Also:
-
GDA_MX_GONIO_ROTATION_AXIS_RECORD
Suffix of epics motor record for gonio rotation axis.- See Also:
-
GDA_MX_GONIO_ROTATION_AXIS_RECORD_DEFAULT
Default value for GDA_MX_GONIO_ROTATION_AXIS_RECORD- See Also:
-
GDA_ANNEAL_IS_SC_INDEPENDENT
Boolean of whether to show the panel created by SharedGuis.createAnnealPanel. For when ann- See Also:
-
GDA_ANNEAL_SCRIPT
- See Also:
-
GDA_ANNEAL_HAS_GEOBRICK_OUT_STATE
The Geobrick anneal out has no state because it has no limit switch- See Also:
-
GDA_MX_SHOW_WAVELENGTH
Whether the wavelength column should be shown in the data collection table. Useful for fixed-wavelength beamlines, for example, where this should befalse.- See Also:
-
GDA_MX_USE_HOLDER_SAMPLE_IN_PREFIX
If the sample changer is available, add _MxSy_ to the prefix- See Also:
-
GDA_SHOW_MOVECRYOJET
- See Also:
-
EIGER_MAX_PREFIX_LENGTH
- See Also:
-
PILATUS_MAX_PREFIX_LENGTH
- See Also:
-
GDA_MX_MINIKAPPA_AVAILABLE
- See Also:
-
GDA_MX_SMARGON_AVAILABLE
- See Also:
-
GDA_MX_HUMIDITYCONTROLLER_AVAILABLE
If the HC1 device is ready and available for use in experiments- See Also:
-
GDA_MX_APERTURE_SCANNABLE
- See Also:
-
GDA_MX_SHOW_SLIT_SIZE
Show the slit gap and controls for showing slit gap in GUI- See Also:
-
GDA_PX_PLATESCREENING
- See Also:
-
GDA_MX_SHOWFOLDER
If set to false then the columns related to data folder will be hidden in the data collection table- See Also:
-
GDA_MX_USINGISPYB
Assumed to be true, but if set to false then parts of the UI relating to sample ids etc. may be switched off- See Also:
-
GDA_MX_SHOWAXISCHOICE
If the users are to be given a choice of which axis to oscillate: Phi or Omega- See Also:
-
GDA_MX_SHOWTWOTHETA
If the detector is mounted on a two theta arm- See Also:
-
GDA_MX_SHOWKAPPANOPHI
Rather than having both kappa and phi displayed by the GDA_MX_MINIKAPPA_AVAILABLE property, only show kappa- See Also:
-
GDA_MX_BEAMSIZEVALUESY
- See Also:
-
GDA_MX_BEAMSIZETABLEUNITS
- See Also:
-
GDA_MX_ISBEAMSIZEXEDITABLE
- See Also:
-
GDA_MX_IS_DETECTOR_DISTANCE_MONITORED
- See Also:
-
GDA_MX_IS_RESOLUTION_READ_ONLY
- See Also:
-
GDA_MX_IS_DETECTOR_DISTANCE_READ_ONLY
- See Also:
-
GDA_MX_ADJUST_RESOLUTION_WITH_ENERGY_CHANGE
- See Also:
-
GDA_MX_BEAMSIZEDROPDOWN
- See Also:
-
GDA_MX_CURRENTBEAMSIZEDEFAULT
- See Also:
-
GDA_MX_SHOWBEAMSIZEX
- See Also:
-
GDA_MX_DC_BEAMSIZEXLABEL
- See Also:
-
GDA_MX_BO_BEAMSIZEXLABEL
- See Also:
-
GDA_MX_BEAMSIZEXUNITS
- See Also:
-
GDA_MX_BEAMSIZEXSCANNABLE
- See Also:
-
GDA_MX_BEAMSIZEXUPDATECMD
- See Also:
-
GDA_MX_SHOWBEAMSIZEY
- See Also:
-
GDA_MX_DC_BEAMSIZEYLABEL
- See Also:
-
GDA_MX_BO_BEAMSIZEYLABEL
- See Also:
-
GDA_MX_BEAMSIZEYUNITS
- See Also:
-
GDA_MX_BEAMSIZEYSCANNABLE
- See Also:
-
GDA_MX_BEAMSIZEYUPDATECMD
- See Also:
-
GDA_MX_SHOW_DOSE_FIELD
- See Also:
-
GDA_MX_MIN_BEAMSIZEX_MICRONS
- See Also:
-
GDA_MX_MIN_BEAMSIZEY_MICRONS
- See Also:
-
GDA_MX_SHOW_XYZ
Include sample centring X/Y/Z positions in data collection parameters- See Also:
-
GDA_MX_SHOW_XYZ_V2
Include V2 sample centring X/Y/Z position tools in data collection GUI (I23-1191)- See Also:
-
GDA_MX_SHOW_EXPERIMENTTYPE
True if data collection form toExperiment Type selection- See Also:
-
GDA_MX_ENABLE_DOSE_OPERATION
True if data collection form to show option to set a target dose- See Also:
-
GDA_MX_ENABLE_DOSE_BY_RESOLUTION_OPERATION
- See Also:
-
GDA_MX_SHOW_USECENTRINGOMEGA
- See Also:
-
GDA_MX_SAMPLECHANGERPRESENT
True if a sample changer is in use- See Also:
-
GDA_MX_ASSOCIATE_CENTRING_WITH_DATA_COLLECTION
- See Also:
-
GDA_PX_GRIDSCAN_ALLOW_COLLECT_USING_COLUMNS
Allow user to use columns while setting up a grid scan- See Also:
-
GDA_PX_GRIDSCAN_ALLOW_COLLECT_USING_ROWS
Allow user to use rows while setting up a grid scan- See Also:
-
GDA_PX_GRIDSCAN_KEEP_BACKLIGHT_IN
- See Also:
-
GDA_PX_FLUORESCENCE_KEEP_DETECTOR_OUT
- See Also:
-
GDA_PX_GRIDSCAN_PMAC_AVAILABLE
Allow fast grid scans using the fast grid scan PMAC program- See Also:
-
GDA_PX_GRIDSCAN_PMAC_USE_SMARGON
- See Also:
-
GDA_PX_GRIDSCAN_BOX_SELECTION
Should the GUI allow grid scan boxes to be selected for data collections?- See Also:
-
GDA_PX_GRIDSCAN_PMAC_FLUO_DATA
Should fluo data be taken during fast grid scans?- See Also:
-
GDA_MX_FLUO_EDGE_FILE
The name of the file storing a list of edges acceptable for fluo scans. This acts as an override for what may be hard-coded in UI classes. Edges should be listed one per line in the format element-edge e.g. Fe-K- See Also:
-
GDA_MX_STANDARD_EXPT_FOLDER
Folder containing Extended Collect Request xml files. If this property is set and valid files exist then items will be added to the Data Collection Table popup menu to allow the user to add new experiments to the table with the values in these files.The labels added to the popup menu is based on the name of the xml files e.g. if a file called Basic_Scan.xml exists then an entry will be added to the popup menu named 'Add Basic Scan'
- See Also:
-
GDA_MX_ANNEAL_NUMBER_OF_TIMES_TO_CHECK_CRYOJET_OUT
During anneal, change the number of times to do the timeout check when the cryojet fine is moved out- See Also:
-
MAX_ANNEAL_TIME
- See Also:
-
NUM_DIGITS_FOR_IMAGE_NUMBER
public static final int NUM_DIGITS_FOR_IMAGE_NUMBERNumber of digits to use for the image number in filenames. -
IMAGE_NUMBER_FORMAT
Format string for the image number component of a filename template. -
MAX_IMAGE_NUMBER
public static final int MAX_IMAGE_NUMBERMaximum possible image number -
NUM_DIGITS_FOR_RUN_NUMBER
public static final int NUM_DIGITS_FOR_RUN_NUMBERNumber of digits to use for the run number in filenames. -
RUN_NUMBER_FORMAT
Format string for the run number component of a filename template. -
NUM_DIGITS_FOR_FOLDER_NUMBER
public static final int NUM_DIGITS_FOR_FOLDER_NUMBERFormat string for the number component of a folder name. -
FOLDER_NUMBER_FORMAT
Format string for the number component of a folder name. -
RUN_NUMBER_FORCE_ASCENDING
public static final boolean RUN_NUMBER_FORCE_ASCENDINGFlag to force ascending run numbers in a collection series when using manually specified run numbers I23-866 -
DO_AUTOPROCESS_EVERY_RUN
public static final boolean DO_AUTOPROCESS_EVERY_RUNFlag to autoprocess incrementally for every run of a data collection group. -
GDA_PX_CORNEASCAN
- See Also:
-
GDA_GTAB_SCANNABLE_NAME
- See Also:
-
GDA_MX_SKIP_FILE_EXISTENCE_CHECK
- See Also:
-
GDA_MX_MIRRORS_OR_CRLS
- See Also:
-
GDA_CLIENTSAMPLEMOVEMENTSERVICE_GONIOONLEFT
used for clientSampleMovementService, gonioOnLeftOfImage field- See Also:
-
GDA_PX_FLUORESCENCESPECTRUM_ALLOWAPERTURESELECTION
- See Also:
-
GDA_MX_MOTD_FILE
Used for Message of the Day (Messages > Beamline notes)- See Also:
-
GDA_MX_SHOW_SCREENING_INSTRUCTIONS
- See Also:
-
GDA_MX_SCREENING_OSCILLATION
Default screening values.- See Also:
-
GDA_MX_SCREENING_DELTA
- See Also:
-
GDA_MX_SCREENING_NUMBER_OF_IMAGES
- See Also:
-
GDA_MX_SCREENING_EXPOSURE_TIME
- See Also:
-
GDA_MX_SCREENING_TRANSMISSION
- See Also:
-
GDA_MX_SCREENING_FIRST_IMAGE_NUMBER
- See Also:
-
GDA_MX_DEFAULT_TOTAL_OSCILLATION
- See Also:
-
GDA_MX_DEFAULT_EXPOSURE_TIME
- See Also:
-
GDA_MX_DATACOLLECTION_SHOW_NEW_GUI
- See Also:
-
GDA_MX_DATACOLLECTION_SHOW_ASTRA_STRATEGY
- See Also:
-
GDA_MX_DATACOLLECTION_ASTRA_STRATEGY_FILE_PATH
- See Also:
-
GDA_MX_DATACOLLECTION_STRATEGY
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_TOTAL_OSC
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_TRANSMISSION
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_SINGLE_START
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_SINGLE_OSC
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_WEDGES_START
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_WEDGES_OSC
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_GPHL_START
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_GPHL_OSC
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_SINGLE_IMAGE
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_WEDGES
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_GPHL
- See Also:
-
GDA_MX_DATACOLLECTION_SCREENING_TABS_SHOW_TARGET_DOSE
- See Also:
-
GDA_MX_DATACOLLECTION_WRITE_TO_JSON_FILE
- See Also:
-
GDA_MX_INSITU_USE_PORTRAIT_DISPLAY
- See Also:
-
GDA_MX_SHUTTER_OPEN_TIME
- See Also:
-
GDA_MX_DETECTORSHUTTER
- See Also:
-
GDA_MX_ENABLE_AUTO_CENTRING
- See Also:
-
GDA_USE_ROI
- See Also:
-
ROI_MODE_XRC_TIMEOUT
public static final int ROI_MODE_XRC_TIMEOUT -
FULL_MODE_XRC_TIMEOUT
public static final int FULL_MODE_XRC_TIMEOUT -
GDA_MX_MULTIPOINT_DATA_COLLECTION_ENABLED
- See Also:
-
GDA_MX_MULTIPOINT_LABEL_SIZE
- See Also:
-
GDA_MX_MULTIPOINT_LABEL_COLOUR
- See Also:
-
GDA_MX_MULTIPOINT_CROSSHAIR_SIZE
- See Also:
-
GDA_MX_MULTIPOINT_CROSSHAIR_COLOUR
- See Also:
-
GDA_MX_CLEAR_COMMENT_VISIBLE
- See Also:
-
GDA_MX_CELAR_COMMENT_SELECTED_BY_DEFAULT
- See Also:
-
GDA_MX_PRE_BEAM_DUMP_MARGIN_SECONDS
- See Also:
-
-
Method Details
-
sampleChangerBlockedPositionsEnabled
public static boolean sampleChangerBlockedPositionsEnabled() -
getMaxSampleHolder
public static int getMaxSampleHolder() -
getMinSampleHolderIndex
public static int getMinSampleHolderIndex() -
getCoordSystemCentringPauseInSeconds
public static double getCoordSystemCentringPauseInSeconds() -
isWavelengthSelectable
public static boolean isWavelengthSelectable() -
getMaxPrefixLength
public static int getMaxPrefixLength(boolean usingEiger) -
isMiniKappaAvailable
public static boolean isMiniKappaAvailable() -
isSmarGonAvailable
public static boolean isSmarGonAvailable() -
isHumidityControllerAvailable
public static boolean isHumidityControllerAvailable() -
isPlateScreeningMode
public static boolean isPlateScreeningMode() -
allowGridScanUsingColumns
public static boolean allowGridScanUsingColumns() -
allowGridScanUsingRows
public static boolean allowGridScanUsingRows() -
keepFluorescenceDetectorOut
public static boolean keepFluorescenceDetectorOut() -
allowFastGridScansUsingPmacProgram
public static boolean allowFastGridScansUsingPmacProgram() -
useSmarGonForFastGridScans
public static boolean useSmarGonForFastGridScans() -
allowGridScanBoxSelection
public static boolean allowGridScanBoxSelection() -
collectFluoDataDuringFastGridScans
public static boolean collectFluoDataDuringFastGridScans() -
isIspybAvailable
public static boolean isIspybAvailable() -
isCorneaScanMode
- Throws:
Exception
-
skipFileExistenceCheck
public static boolean skipFileExistenceCheck() -
usingCrlsOnly
public static boolean usingCrlsOnly() -
usingMirrorsOnly
public static boolean usingMirrorsOnly() -
showApertureInFluoSpectrumParameters
public static boolean showApertureInFluoSpectrumParameters() -
areLineScansAvailable
public static boolean areLineScansAvailable() -
areHelicalLineScansAvailable
public static boolean areHelicalLineScansAvailable() -
areFluorescenceScansAvailable
public static boolean areFluorescenceScansAvailable() -
isSampleHandlingAvailable
public static boolean isSampleHandlingAvailable() -
getShutterOpenTime
public static double getShutterOpenTime() -
hasDetectorShutter
public static boolean hasDetectorShutter() -
isUsingRoi
public static boolean isUsingRoi() -
getTimeoutForCentringResults
public static int getTimeoutForCentringResults() -
isMultipointDataCollectionEnabled
public static boolean isMultipointDataCollectionEnabled() -
isClearCommentVisible
public static boolean isClearCommentVisible() -
isClearCommentSelectedByDefault
public static boolean isClearCommentSelectedByDefault() -
getSEMDriveMapping
-
getPreBeamDumpMarginInSeconds
public static int getPreBeamDumpMarginInSeconds()
-